Wed Feb 9 08:01:41 2022, OH, CB, G-HC, Results from no beam overnight 
|
08:52 Overnight the DAQ was left running to check if there was any individual radioactivity in the Si.
This is stored in R7_75
|
Thu Jul 13 08:46:44 2023, TD, Replacing wire seal 
|
09.30 CARME chamber status
|
Mon Sep 25 16:33:32 2023, TD, Removed tape from heaters TC  
|
17.33 IE514 pressure 6.4e-11 mbar, ambient temperature +20.8 deg C
|
Wed Jan 17 10:31:05 2024, TD, Remote control of CAEN N1419ET from nnrpi1
|
Login to nnrpi1 from carme-gsi using ssh
|
Thu Jan 27 14:04:34 2022, OH, JM, Remote access to CARME server
|
1. ssh into lxpool network: ssh <user>@lxpool.gsi.de
2. ssh into atppc022.gsi.de: ssh carme@atppc022.gsi.de (This is an rpi access server. Don't run code)
3. ssh into carme server: ssh npg@carme-gsi
|
Thu Feb 24 07:25:53 2022, OH (Remote), Recover of system after crash
|
08:00 Woke up to message from Carlo explaining the server had crashed.
Anydesked into the machine.
Merger was showing just data transfer it was not showing Transfer => Merger
|
Mon Jun 16 22:06:00 2025, CB, AR, EW, RUN 90-91 14x
|
23:00 DAQ restored and data now being written okay. Checks all okay, see screenshots below. Histograms reset. 40 bars in bottle @ 23:45 - should last
30 hours.
12:00 Checks all ok. No problems with aidas or writing to disk. |
Tue Nov 24 12:24:01 2020, JM, RGA scans
|
RGA laptop passcode: tw%11&STA-
RGA scans set up and plotting data as of 11:44
RGA is set to the multiplier mode (4.44e-8mbar) with an accuracy setting of 8, these are for the most sensitive scans. RGA is set to scan masses |
Thu Jun 13 14:46:17 2024, TD, R66 beam overlap test
|
15.45 Start data file G22-201/R66
Beam/target overlap test |
Thu May 22 06:41:10 2025, JG, TD, R6 - R11 scrapers test / pulser walk 33x
|
07.00
JG found target off (target data stopped c. 02.30 this morning) and beam off (not re-enabled from previous CRYRNG area entry - target related?) JG re-enables |
Mon Jun 10 05:24:49 2024, TD, RSS, DB, R57 16x
|
06:45 New data file G22-201/R57
Right arm at 64.60 mm
|
Sat Jun 8 14:49:14 2024, TD, R51 - contd 30x
|
15.40 DSSSD bias & leakage current OK - attachments 1-2
log shows variations due to ambient temperature changes in CRYRING hall
|
Tue Feb 20 13:20:44 2024, OF, AG, JG, ML, R48 status check   
|
Doing afternoon check. Checked the leakage currents, all ok. Attachment 1
Temperatures ok, attchment 2
System wide check, all correct. |
Wed May 21 14:38:22 2025, JM, TD, R4 and R5 - background 7x
|
15:30
Waiting for the target to cool down. ETA is 16:30 for cooling to be finished and then we need to try and make target.
While waiting we havve been running a background (R3). I noticed the rates on aida10 had increased (300k) and aida09 were slowlly creeping up |
Wed Sep 16 17:46:36 2020, JM, Pumping of Chamber
|
A pump cart and leak cart have been connected to the chamber via the top turbo at the back of the chamber. An air line has also been connected to the
valve to open/close it. The pirani gaige has also been replaced by a blank flange, a gauge on the pump cart can measure the pressure in its range. The
turbo fan and controller were set up and pumping was started after lunch. |
Tue Jul 18 08:16:28 2023, TD, Pumping down. Issues found with one detector.
|
09.15 Following overnight pumping of CARME chamber by Leybold turbo
Pirani 1.00e-3 mbar IE514 2.83e-7 mbar
|
Sun Jun 25 10:08:30 2023, JM, TD, Pump-down   
|
23/06/23
Pumpdown started with the scroll pump 17:15 . The turbo-pump was started 17:55. Gate valve was closed and the turbo turned of at 19:20 due to
no working fan on the turbo. |
Mon Jul 24 10:40:14 2023, RSS, TD, Pump down continues 
|
11.40 Pirani 1.00e-3 mbar IE514 2.23e-08mbar
15.15 7d/24h vacuum plots |
Wed Jul 5 19:27:13 2023, JM, Pump down and leak test  
|
CARME and the interaction chamber have been pumping for ~26 hours, P=4.15e-7 mbar, attached figures show the current pump down plot in comparison to
the two previous pumpdowns in the last 2 weeks and the pump down with the bare chamber in 2020. The gradient of the pressure reduction appears to be improving
for each pump down, either due to conditioning of the chamber or the reduction in leaks. |
Tue Feb 15 14:59:37 2022, JM, CB, RJC, Pulser walkthrough 11x
|
16:00
Check Fee temperatures: all green (see attachment 1)
Check statistics tab: ADC data items (see attachment 2) |