Entry  Tue Feb 22 10:06:50 2022, RSS, OF, ML, PMT channel 
YR00DA1ZB is the channel for PMT rate.
Entry  Fri May 31 21:53:20 2024, CB EM, JM,EM AB, Overnight beam tests 17x
22:53

Run 15 started

Beam tests. We increased the storage time to 2min and switched to bunched beam so that we can see the beam lifetime over the cycle.
Entry  Wed Oct 6 09:39:59 2021, PB CB, Original dirty assembly of CARME internals 7x
PB found pictures of the original dirty assembly, before screws and components were shipped to DL.
See attached.
Entry  Wed Oct 13 15:16:12 2021, TD, Orientation of 2x 78 way D connector feedthroughs 20211013_152518.jpg20211013_152529.jpg20211013_152549.jpg20211013_152606.jpg

Looking downstream
Entry  Sun Apr 6 11:40:48 2025, TD, Options DB reconstructed Screenshot_from_2025-04-06_12-59-39.png
Options directory corrupted unable to change Options settings per https://elog.ph.ed.ac.uk/CARME/664

Move old Options directory to Options.BAK-060425
Entry  Wed Feb 23 16:57:35 2022, CB, CN, Optimising motor motion 
Optimised motor motion during run R58

Moved 2 mm in (20.3 potentiometer)

29.47 potentiometer
Entry  Wed Jan 31 09:49:38 2024, JM, TD, Optimising - Wednesday 31 25x
Removed ground from DSSD #1 to the chamber and set MACBs so that timestamp is from the NIM WR emulator not the VME VETAR2.

Root MACB mode 0x3 -> 0xd - all other MACB modes 0x3 ( unchanged ).

No timestamps observed from the emulator. Re-set to take timestamps from the VETAR" unit -> all timestamps observed.
Entry  Thu May 30 06:00:51 2024, Yuri Nikos ..., Open the cave 
Nikos needs to enter the cave to work on the target cooling.

 
Entry  Tue May 13 15:07:42 2025, JM, Online monitoring code Screenshot_from_2025-05-13_15-14-12.pngScreenshot_from_2025-05-13_15-14-38.pngScreenshot_from_2025-05-13_15-28-48.png
CARME_Sort histogramming has been updated to aid our understanding online.

> XY plots have been updated to reflect the true position of the beam relative to each detector. Online we will be able to layout in a grid
to reflect looking downstream in YR09 (see attachment 1)
Entry  Tue Jul 2 17:43:32 2024, TD, Offline analysis G22-201/R89 9x
Offline analysis data file G22-201/R89_0 to R89_13 (c. 6h40m data)

DSSSD#1 & 3 36.54mm -> ~ 2.3-8.6 deg 
Entry  Sun Jun 2 14:30:16 2024, TD, Offline analysis G22-201/R22_0 - R22_11 8x
Offline analysis data file G22-201/R22_0 to R22_11

Sort program variables https://elog.ph.ed.ac.uk/CARME/240530_180659/variables.dat
Entry  Sat Jun 1 16:38:13 2024, TD, Offline analysis G22-201/R17 8x
Offline analysis data file G22-201/R17

Sort program variables https://elog.ph.ed.ac.uk/CARME/240530_180659/variables.dat
Entry  Sat Jun 1 13:33:50 2024, TD, Offline analysis APR24/R3 (alpha background)  9x

Offline analysis data file APR24/R3 (alpha background)
Entry  Wed May 29 17:42:06 2024, TD, Offline analysis APR24/R2 (pulser walkthrough) 17x
ADC offsets using pulser walkthrough data from data file APR24/R2

ch = channel + ( module * 64 ) + ( range * 2048 )
Entry  Sun Nov 28 15:57:45 2021, CB (remotely), Nov 2021 temperatures 211128164424_temperatures.zip
Attached
Entry  Sun Jun 25 09:41:06 2023, JM, TD, Normal turbo-pump operation 6x
Attached are images showing the messages for normal turbo-pump operation
Entry  Tue Feb 15 21:28:14 2022, JM, Noisy Channels 2022-02-15_22-17-51-aida1-3*L.png2022-02-15_22-23-12-aida1-4-stat-log.png
All fees and asics checked to find noisy channels. Channels are listed along with the noise peak height for the low energy plot at 22:00. Run started
~17:45. 

Attachment 1 shows 3*L for aida01 with x axis zoomed in on noise peak. Attachment 2 shows stat for aida01-04 in log scale. 
Entry  Sat Aug 13 10:23:59 2022, JM, TD, Noise tests pt3 and pulser walk 15x
Waveforms and discriminator ON. Pulser re-connected to all fees via daisy chain. V=100v, I=8.36uA

 

Pulser peak widths (ch FWHM)
Entry  Fri Aug 12 09:06:57 2022, TD, JM, Noise tests pt2 30x

10.07 DSSSD bias -100V leakage current -8.65uA
Entry  Thu Aug 11 09:25:13 2022, JM, TD, Noise tests 14x

Moved pneumatic motor in and out several revolutions

Biased detector up to 100 V, I ~ 12.4 uA ambient temp c. 21 deg C