00.00 Degrader changed to 4mm + some of variable (need to confirm)
Extra 1mm removed.
00.30 R63
95Br BigRIPS setting -> 94Se
Degrader ?
FEE64 nnaida1-16 stat spectra (attachment 1)
00.50 Beam tuning for 100Kr started
Degrader changed to 4mm fixed + 0.6mm + 0.2mm variable
=> total 4.8mm
09.40 Found detector bias DSSSD #1 tripped
Detector trips at +20uA at ~ +10V
10.45 R63 stopped
10.45 Bias(V) I_L(uA) set 50uA trip current
10 45.105
20 trips at ~15V
2 4.450
4 ~7.4 rising to ~9.4 over ~2 mins
6 20.4 and rising
Not open circuit. Not Ohmic. Radiation damage?
Time between successive 32000 block data files changes from ~5-6 mins to ~3-4 mins
07.57-08.00. SEASTAR Elog records F/C drops in at 08.20. Nishimura-san reports MAKI
DAQ stopped/stalled c.08.00. Abrupt failure?
11.15 R64
DSSSD #1 off
nnaida6, 10, 11 & 16 slow comparator -> 0xff, medium/low energy selection -> 0x1
12.10 R64 stopped
beam off
12.25 R65
14.09 R65 ends
R66
New ZDS setting (change in B-rho)
F11 rate ~4kHz
~15.00 Degrader changed -> 4mm + 1.3mm (variable) Al
16.01 R66 ends
17.00 Checked CAEN N1419 ch #1 with/without bias cable and disconnecetd from DSSSD #1 - zero current
at bias => problem forward of adaptor PCB
17.28 R68
100Kr setting
While tuning ZDS, observe few, if any, implants in DSSSD #2 & #3
18.30 Degrader changed to 4mm fixed + 0.4mm variable
19.16 R68 ends
R69
Degrader -> 4mm + 0.4mm (variable) Al |