<p>Powered on system ok.</p>
<p>Electrical connection between crate and mezzanines still persists, but only to the top cool plate (no connection to the bottom or sides), eith resistance ~5-10 ohms</p>
<p>Seems to have no effect on performance though as without the kaptons connected, still see pulser widths of ~14ch at t_shape = 8us.</p>
<p>Connecting cables reinstates double peaking in n-side.</p>
<p>Without HV or detector connected, nnaida13+14 gibe single Gaussian peaks of width ~45ch, with almost perfect waveforms.</p>
<p> - Patrick saw this too, I'm not going crazy, but on zeroing histograms (didn't change a single other thing) it went back to having double peaks again.</p>
<p>Replaced nnaida13 shielded kapton with unshielded one to see if that is causing this effect</p>
<p> - still see double peaking but width of tallest peak is down by ~25% (likely just due to capacitance)</p>
<p>With shielding on, do see much noisier waveforms however.</p>
<p> - note: attachments 1 and 2 were taken using exactly the same threshold of 512. When shielded, bumping this up to 1000 still made little difference and 'actual' traces were only caught ~25% of the time</p>
<p>Changed firmware to V8.</p>
<p>No difference in performance.</p>