13/12/15
Current jumper setup:
- detector 2977-15 connected to nnaida1,3,5,8 with GND kapton to nnaida1
- bias of +75V to nnaida3 and nnaida5
- Jumpers:
- nnaida1 > Lk1. If Lk3 3 connected unable to bias detector
- nnaida3 > Lk3. if Lk1 connected unable to bias detector as the HV ground reference is connected at both p- and n-sides of detector.
- nnaida5 > Lk1 + Lk3
- nnaida8 > Lk3
Powered system up as it was last night -> 2977-15 connected to nnaida1,3,5,8 with only 4 FEE modules powered.
- n-side widths ~190-200 ch
- p-side widths ~100ch
- jumpers as above, ASIC setting as 2015Dec12_smallSetup, wav thresholds all at 500 apart from nnaida5 @ 250.
Spend remainder of day trying to fit mezzanine cables.
14/12/15
Completed installation of mezzanine cables.
Resistance between all mezzanines and cooling crates still ~200 kOhms (up from ~2-5 Ohms on our arrival), apart from nnaida7+8, which is now at ~5 Ohms
However, all see resistance to main vertical copper pipe of ~3 Ohms.
As best as I can see, isolating tape in still in place for all FEEs => unknown point of contact.
> LEMO cables to test inputs carry ground connection from NIM bin -> NIM bin in electrical contact with support frame -> coolant loop in electrical contact with support frame => electrical connection still present between FEEs and coolant loop, even in insulating guide pins are introduced
> this problem is not observed at DL because the NIM bin in on the wooden table top and isolated from the coolant loop and any support structure
> is it possible to isolate the NIM bins from the support structure?
> need to come up with some more systematic and fixed way/positions for measuring these resistances, i.e. always measure mezzanine to X.
Resistance between all mezzanines now ~ 0.5 Ohms.
n-side widths ~ 120ch and p-side widths ~ 120ch => peak widths almost halved in n-side, but increased by ~20% in p-side.
Attempted to bias previously tripped detector.
- same jumper combination as other detector caused failure to bias > leakage current jumps to 50uA at ~2V and trips shortly after, ramping back down from there.
- removed all jumpers and see same thing.
Will try again in the morning by attaching 3131-11 to cables currently on 2977-15, but it looks like 3131-11 really is defective somehow. |