18.00 Swapped input cables to nnaida11 and nnaida14 both cables LH coupler + 2x 34-way Samtec ribbon cables + 3M 1245 1.4mil Copper screen w/ground wire RH coupler not fitted both ground wires tied to BNC cable from pulser to AIDA adaptor PCBs at BNC(F)-Lemo-00(M) adaptor Before swap, input cable attached to nnaida11 exhibited typical peak widths ~50 ch FWHM, cable attached to nnaida exhibited double peaking and other structure This is unchanged following swap => grounding and/or AIDA adaptor PCB significant *not* cable See attachments 1-4 post swap nnaida11 ASIC 2*L ~40-60 ch FWHM pulser peak width standard ASIC settings, shaping time 2us