<p>I have changed the T9 system to be using Version 8.</p>
<p>Tests using the detector and pulser seem to be ok.</p>
<p>Transmission to the Merge and Merging operating ok</p>
<p>Shaping time 8uS and threshold 64.</p>
<p>Note:-</p>
<p style="margin-left: 40px;">The Leading Edge Discriminator has changed in Version 8 and is a true implementation of a Leading Edge Discriminator. With hysteresis which requires the input signal to go below the threshold before allowing another trigger.</p>
<p style="margin-left: 40px;">The Delay setting on the LED web page is removed.</p>
<p style="margin-left: 40px;">The LED web page is aware of the version being used ( shown at the top of the page ) as the Threshold for the LED is an ADC value ( 0 to 16383 ) rather than a difference from the current baseline.</p>
<p style="margin-left: 40px;">For a positive waveform where the baseline is expected to be about 8192 a value of 9000 will potentially give a trigger.</p>
<p style="margin-left: 40px;">For a negative waveform where the baseline is expected to be about 8192 a value of 7000 will potentially give a trigger.</p>
<p style="margin-left: 40px;">Example window attached.</p>
<p>I would value feedback.</p>
<p> </p> |