The attached VHDL file should be saved to /MIDAS/Aida and the FlashPgm.csh file edited to load "FEE_GF_Feb16_16.bin" Use the Run Control Expert command to update the firmware on all the FEE64s. Power cycle the FEE64s. This will give a 25Mhz clock from the MACB Fast NIM output 0. The attached .js and .tml files will change the Control browser window to allow access to a Correlation control window. I have already installed the Correlation directory and files on aidas. The Correlation Control window has two extra buttons. One to cause a Correlation trigger and the other to cause a reset to be sent out. The Reset is only sent out if the Run Control is in the GO state. To complete the change first make backup copies of the Aida.js and Aida.tml files that are currently in /MIDAS/TclHttpd/Html/AIDA then copy the attached new versions into /MIDAS/TclHttpd/Html/AIDA. These two files are in the attached .zip file. The Elog didn't like the .js file extension. The Browser window reload should work but it may be required to clean your browser history to ensure the new version of the file is used.
Attachment 1: FEE_GF_Feb16_16.bin  3.222 MB
Attachment 2: Aida_js_tml_May16.zip  3 kB
ELOG V3.1.4-unknown