Entry time:
Sun Mar 30 09:19:21 2025
Author
*
:
Subject
*
:
> The firmware of all FEEs was updated to the version in FEE_GF_Feb16_16.bin (https://elog.ph.ed.ac.uk/AIDA/215). > This provides a 25 MHz clock output from MACB, plus a manual reset request signal. > > Attached are screen captures from smooth updating procedure. > > Checked correlation request signal from output2 of MACB: it's a pulse with ~180ns width and -0.7V. We could only > observe it when selecting 'act no all FEEs' in Correlation Control browser window (looking at MACB module on the > top of the tree). However, we didn't check that the output is produced from only one MACB module, if we select > the one of the corresponding FEE.