Jan 20, 2015
Test settings: DSSD biased at 100V, with 3.5 uA leakage current. Pulser at 0.5 V, split and inverted for both
positive and negative polarity. Frequency at 100 Hz.
++SHAPING TIME++
We varied shaping time (Tsh), and adjusted hold time (Th) accordingly. We measured the width of pulser spectra
for two channels using 'integrate' function of spectrum browser.
Tsh Th nnaida#11-2.6L nnaida#13-2.6L
[usec] [usec] FWHM [ch] FWHM [ch]
0.5 2 146 109
1 2 133 113
2 2 131 116
4 6 130 126
8 8 145 136
note: Tsh= ( 0.5*offset9 + 0.5) usec, Th= 2*offset7 usec
The first two data points where saved to file R44_0.gz, and R45_0.gz. Starting from the Tsh=2usec measurement,
NNAIDA#12 was causing communication issues with the merger/tape server, so we don't have files for last few.
Due to limited storange in AIDA PC, data was stored to external drive provided by Vic
(/media/data/TapeData/Test100/), and backued up temporarly also in Edinburhg
(/Disk/ds-sopa-personal/aestrade/data/AIDA/).
++THRESHOLDS++
Next step was to adjust threshold of the slow comparator, in order to lower them and take data looking for
background alpha radioactivity.
At this point NNAIDA#13 had stopped producing data (probably similar issue as NNAIDA#12, which could be
recovered with power cycling FEEs), so we used nnaida#14
FEE Eth rate/channel integral(pulse peak):integral(noise peak)
nnaida#11 0x40 115 Hz (~only pulser)
nnaida#14 0x40 115 Hz(~only pulser)
nnaida#14 0x20 ~350 Hz 2:1
nnaida#11 0x10 ~150 Hz 1:2 (one channel very noisy @ 15kHz)
nnaida#14 0x10 ~5 kHz (asic 1-3) 80:1
0x10 ~3kHz-100 Hz (asic 4)
For background run (see next elog entry) threshold values set at:
nnaida#11= 0x16
nnaida#12= 0x40
nnaida#13= 0x20
nnaida#14= 0x16 |