AIDA |
 |
|
Message ID: 222
Entry time: Thu May 12 15:50:48 2016
|
Author: |
Patrick |
Subject: |
System Checks and firmware update |
|
|
Added two checks to the System wide checks.
Modified the Firmware so the checks work properly.
One check reads the status of the FEE64 clocks and checks against expected value.
The other check reads the value from all the FEE64 Sync Error Counters, pauses for one second, then reads them again and compares with the first reading. Any change is reported as an error.
The Sync Error counter is incremented if the SYNC pulse arrives at a FEE64 when the Timestamp counter in the FEE64 is not at the correct value. |
|
|