|
AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
|
AIDA |
|
|
Message ID: 262
Entry time: Tue May 31 08:06:43 2016
|
Author: |
TD, AE & CG |
Subject: |
Tuesday 31 May 2016 |
|
|
Attachment 3 - AIDA discriminator with pulser ~5Hz
standard ASIC settings (tau=8us, slow comparator 0x10, LEC/MEC comparator 0x9)
17.30 Added jumpers to the front/rear field plate connections (Lk 2/4 + 6/8) on all FEEs corresponding to DSSDs
1 and 6
Saw significant reduction in size of noise peak in waveforms as well as in the fast discriminator rate
from all FEEs.
18.50 R13 (AIDA) + BiCocal_run0012.lmd (EURICA)
207Bi beta-gamma correlation test
Standard ASIC settings (tau=8us, slow comparator 0x10, LEC/MEC comparator 0x9)
BNC PB-4 pulser OFF
Settings DB key 2016May31-18.55.42
Attachment 4 - statistics AIDA discriminators
Attachments 6-7 - time correlation for 207Bi beta-gamma coincidences between EURICA Ge Or and
DSSSD#1 trigger ( .OR.X strips .AND. OR.Y strips ) |
|
|
|
|
|
|
|
|
|
|
|
|
|