Entry time:
Wed Jun 1 09:00:13 2016
Author
*
:
Subject
*
:
> 09.30 Detector biases and leakage currents - ambient temp +24.9 deg C - attachments 1-2 > > 10.08 Connection lost to nnaida27 and nnaida28 > Most recent save of settings DB key 2016May31-23.32.46 > > 12.00 Replaced nnaida17 & nnaida18 with FEE module previously used for nnaida31 & nnaida32 > > Pulser walkthrough > > BNC PB-4 settings > > 90,000 - 10,000 @ 10,000 step > x5 attenuator (was x10 previously) > t_r = 50ns tau_d = 50us > polarity - > rate ~ 150Hz > INT, delay MIN > > polarity + by EG&G Ortec 433A > > N.B. - polarity terminated at end of AIDA adaptor chain, + polarity not terminated > > Spectra saved 12.32 ( to directories tagged around June 1, 2016 05.29 ) > > BNC PB4 pulser rate set to ~5Hz > > 207Bi source at front of snout and 2x 60Co sources above DSSD3 > > 14.00 Rebooted aidas1 to cleanup merge > > Changed config to 24 FEE64s > > Clock issues with nnaida23 - clock not locked status 0x6 - see attachments 3-9 Check the Local Controls for the setting of the LMK3200 control rgister. If you recall from the problem with nnaida5 the value should be 0xB. If the clock still doesn't lock then maybe it's an HDMI cable problem like nnaida5. Since the other FEE64 are ok then it is likely to be just the MACB to nnaida23.