Message ID: 269     Entry time: Wed Jun 1 13:58:48 2016     In reply to: 267
Author: TD - Reply PJCS 
Subject: Wednesday 1 June 2016 
> Check the Local Controls for the setting of the LMK3200 control rgister.
> If you recall from the problem with nnaida5 the value should be 0xB.
> If the clock still doesn't lock then maybe it's an HDMI cable problem like nnaida5.
> Since the other FEE64 are ok then it is likely to be just the MACB to nnaida23.

I attach SYNC stats for WAVE and AIDA - you will notice that most (but not all) FEE64s
produce WAVE SYNCS

Plus (load balanced) statistics for disc, good events