AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
AIDA
Draft saved at 00:00:00
Fields marked with
*
are required
Entry time:
Sat Mar 29 11:35:41 2025
Author
*
:
Subject
*
:
<p>Due to sometimes operating with very noisy systems, both at Daresbury and RIKEN, I have installed a change to the startup of the system.</p> <p>The Runcontrol GO button now only starts the output of SYNC pulses. A further button labelled "Start ASIC Readout" ( see attached photo ) needs to be pressed to start the ASIC ADC, Discriminator and Correlation readout.</p> <p>After carrying out the normal operations to start the system and pressing the GO button the Merger can be checked to be certain it is merging the events. Since there are only SYNC pulses it should soon be apparent if this is not happening and appropriate actions can be taken. ( Kill/Reload the Link and Merger processes using the "Merger for AIDA" icon at the top of the screen is one way. )</p> <p>Once it is certain the merging is in progress then the new button labelled "Start ASIC Readout" ( see attached photo ) needs to be pressed to start the ASIC ADC, Discriminator and Correlation readout.</p> <p>This must be done after every GO.</p>
Encoding
:
HTML
ELCode
plain
Suppress Email notification
Attachment 1:
Drop attachments here...
Draft saved at 00:00:00
ELOG V3.1.4-unknown