09.50 DAQ found stopped No error messages merge/tapeserver - see attachments 1-2 Most recent data file R7_107 last written to 08.13 23.8.16 10.00 Electrical power to AIDA is off for reasons unknown (power saving/maintenance/trip) 3-ph circuit to Julabo FL1106 recirculating chiller is unaffected 10.20 Analysis of R7_107 (most recent data file) shows data synchronised OK 11.30 Yoshida-san informs us that the experimental power to F11 is off for annual inspection/maintenance. Power should be restored this evening, or possibly, sometime tomorrow. 19.00 Power restored FEE64 power *off* - as expected following power outage Detector bias *off* - as expected following power outage Startup AIDA System wide checks, FEE64 temperatures and DSSSD bias/leakage currents - see attachments 3-11 19.24 R8 starts slow comparator 0x10 LEC/MEC fast comparator 0xff ASIC settings 2016Aug20-15.34.53 Discriminator masks -> 0x0 BNC PB-4 Pulser Amplitude 90,000 x5 attenuator IN t_r 50ns tau_d 50us - polarity INT ref, delay MIN, tail pulse 19.31 AIDA SYNC, PAUSE, RESUME, SYNCtime warp, Good data etc statistics - see attachments 12-18