AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
AIDA
Draft saved at 00:00:00
Fields marked with
*
are required
Entry time:
Thu Oct 31 23:06:55 2024
Author
*
:
Subject
*
:
> 11.40 R9 continues > > ASIC settings 2016Aug20-15.34.53 > Discriminator masks -> 0x0 > From Correlation Status & Control tab change Trigger delay from 0x180 to 0x5 > > BNC PB-4 Pulser > Amplitude 90,000 > x5 attenuator IN > t_r 50ns tau_d 50us > - polarity > INT ref, delay MIN, tail pulse > External (random) trigger from EG&G Ortec 462 Start output (Range 81.92us) - frequency c. 120Hz > > B2F/F11 temp +25.2 deg C, d.p. +14.5 deg C, RH 50.4% > > DSSSD bias/leakage current OK - see attachments 1-2 > FEE64 temperatures OK - see attachment 3 > > Some error messages observed for merger - see attachment 4-6 > All related to Q5 (nnaida6) > > System-wide checks OK - see attachments 7-12 > > AIDA SYNC, PAUSE, RESUME, Correlation L, SYNC time warp, Good Events, nnaida6, nnaida5 & nnaida20 > statistics - see attachments > 13-20 > > Analysis of R9_213 shows data synchronised - see attachment 21 > > FEE # versus timestamp (0.1ms / channel) > Note nnaida19 not currently configured
Encoding
:
HTML
ELCode
plain
Suppress Email notification
Attachment 1:
Drop attachments here...
Draft saved at 00:00:00
ELOG V3.1.4-unknown