Entry time:
Sat Mar 29 11:35:40 2025
Author
*
:
Subject
*
:
10.11 R9 continues ASIC settings 2016Aug20-15.34.53 Discriminator masks -> 0x0 From Correlation Status & Control tab change Trigger delay from 0x180 to 0x5 BNC PB-4 Pulser Amplitude 90,000 x5 attenuator IN t_r 50ns tau_d 50us - polarity INT ref, delay MIN, tail pulse External (random) trigger from EG&G Ortec 462 Start output (Range 81.92us) - frequency c. 120Hz B2F/F11 temp +24.8 deg C, d.p. +12.9 deg C, RH 46.8% DSSSD bias/leakage current OK - see attachments 1-2 FEE64 temperatures OK - see attachment 3 Some error messages observed for merger - see attachment 4-12 All related to Q5 (nnaida6) System-wide checks OK - see attachments 13-18 AIDA SYNC, PAUSE, RESUME, Correlation L, SYNC time warp, Good Events, nnaida6, nnaida5 & nnaida20 statistics - see attachments 19-28 Analysis of R9_561 shows data synchronised - see attachmen FEE # versus timestamp (0.1ms / channel) Note nnaida19 not currently configured 10.49 R9 stops