Entry time:
Sat Mar 29 11:18:03 2025
Author
*
:
Subject
*
:
> 10.11 R9 continues > > ASIC settings 2016Aug20-15.34.53 > Discriminator masks -> 0x0 > From Correlation Status & Control tab change Trigger delay from 0x180 to 0x5 > > BNC PB-4 Pulser > Amplitude 90,000 > x5 attenuator IN > t_r 50ns tau_d 50us > - polarity > INT ref, delay MIN, tail pulse > External (random) trigger from EG&G Ortec 462 Start output (Range 81.92us) - frequency c. 120Hz > > B2F/F11 temp +24.8 deg C, d.p. +12.9 deg C, RH 46.8% > > DSSSD bias/leakage current OK - see attachments 1-2 > FEE64 temperatures OK - see attachment 3 > > Some error messages observed for merger - see attachment 4-12 > All related to Q5 (nnaida6) > > System-wide checks OK - see attachments 13-18 > > AIDA SYNC, PAUSE, RESUME, Correlation L, SYNC time warp, Good Events, nnaida6, nnaida5 & nnaida20 > statistics - see attachments > 19-28 > > Analysis of R9_561 shows data synchronised - see attachmen > > FEE # versus timestamp (0.1ms / channel) > Note nnaida19 not currently configured > > 10.49 R9 stops