> 11.00 Moved detector stack forward several cm. 1st DSSD now ~6cm from end of snout. Middle of stack now ~47cm from > base -> approx centre of BRIKEN. > > 4 new 79cm ribbon cables (+kaptons) added to p-side of first two DSSDs. > > Bias and pulser re-connected to all relevant FEEs. > > Detector order unchanged, i.e. DSSD1 has same serial number as in June 2016 expt. Same for DSSDs 2-6. > > DSSD-to-FEE map also unchanged. > > 15.45 R7 starts > > ASIC settings 2016Sep02-18.17.01 > > BNC PB-4 > Amplitude 90,000, x2 attenuator IN > Frequency c. 26Hz > > Master MACB > mode 5 > 50MHz external clock (rear panel) > BNC PB-4 trigger output -> Ortec SCA -> LeCroy 429A -> Master MACB Correlation DAQ Accepted Trigger - not > working - firmware issue? > > DSSSD bias/leakage current OK - see attachments 1-2 > FEE64 temperatures OK - see attachment 3 > > System-wide checks OK *except* nnaida19 fails ASIC timestamp check (no change) *and* nnaida7, nnaida21 > fail ADC calibration > > AIDA SYNC, PAUSE, RESUME, Correlation L, SYNC time warp, Good Events, AIDA disc, nnaida6, nnaida5, > nnaida19 & nnaida20 statistics - see attachments 4-14 > > Analysis of R7_0 - ADC data synchronised