AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
|
AIDA |
 |
|
Message ID: 368
Entry time: Sun Oct 2 06:12:54 2016
|
Author: |
TD |
Subject: |
Sunday 2 October |
|
|
14.15 DSSSD bias/leakage current OK - see attachments 1-2
FEE64 temperatures OK - see attachment 3
B2F/F11 temperature +27.5 deg C, d.p. +10.6 deg C, 34.3 RH %
14.24 R8 starts
ASIC settings 2016Sep02-18.17.01
Master MACB serial #13 (see https://elog.ph.ed.ac.uk/AIDA/367)
Master MACB mode 5
NIM 50MHz -> Phillips 726 NIM-TTL-NIM -> Master MACB - external clock (rear panel)
Note Fuji Diamond Tool Ltd 1020 NIM-TTL level adaptor (NIM, pink front panel) with signal
rise and falltimes quoted as <10ns produced poor quality 25/50MHz clock - the Phillips 726
has significantly faster rise and falltimes and drices a higher amplitude signal to 50R load.
BNC PB-4 trigger output -> Ortec SCA -> LeCroy 429A (500ns width) -> Master MACB - Correlation DAQ
Accepted Trigger
BNC PB-4 frequency c. 26Hz
Statistics - see attachments 4-14
Master Timestamp - see attachment 15
Correlation Status & Control - see attachment 16
Successive correlation scaler events from R8_0
*** MBS information: block: 120 ptr: 7827 data: 0x868068B4 module: 6 information type: 8 information
index: 0 MBS data:68B4 ts: 0x00000019979ADAF8
*** MBS information: block: 120 ptr: 7829 data: 0x86812F00 module: 6 information type: 8 information
index: 1 MBS data:2F00 ts: 0x00000019979ADAF8
*** MBS information: block: 120 ptr: 7831 data: 0x86820000 module: 6 information type: 8 information
index: 2 MBS data:0000 ts: 0x00000019979ADAF8
*** MBS information: block: 126 ptr: 7005 data: 0x8680FDF4 module: 6 information type: 8 information
index: 0 MBS data:FDF4 ts: 0x0000001997D52FF8
*** MBS information: block: 126 ptr: 7007 data: 0x86812F0E module: 6 information type: 8 information
index: 1 MBS data:2F0E ts: 0x0000001997D52FF8
*** MBS information: block: 126 ptr: 7009 data: 0x86820000 module: 6 information type: 8 information
index: 2 MBS data:0000 ts: 0x0000001997D52FF8
0x2F0EFDF4 - 0x2F0068B4 = 0xE9540 = 955712 * 40ns (assume 25MHz) = 38228480ns = 38.2ms
0x1997D52FF8 - 0x19979ADAF8 = 0x3A5500 = 3822848 * 10ns = 38228480ns = 38.2ms
=> correlation clock frequency = 25MHz
Master MACB (serial #13) mode = 5, 50MHz clock to external clock input (rear panel)
clock output 25MHz
reset output reset not observed at DAQ GO
reset observed using Correlation status & control tab -> Force
Correlation Reset
reset request output not tested
correlation DAQ accepted trigger output c.26Hz
Could not persuade MACB serial #13 to operate in mode 0, external clock input (rear panel) disconnected)
Persistent failures to ReSYNC by nnaida11, 12, 17, 18, 19, & 20 (see https://elog.ph.ed.ac.uk/AIDA/357)
following RESET/SETUP, FEE64 power cycle, MACB NIM bin power cycle and re-seating master MACB HDMI cables
Switch back to previous Master MACB serial #11
Master MACB (serial #11) mode = 0
clock output 25MHz
reset output reset not observed at DAQ GO
reset not observed using Correlation status & control tab -> Force
Correlation Reset
reset request output not tested
correlation DAQ accepted trigger output not observed |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|