10.29 DAQ left running overnight Histograms only - no data transfer to Merger & TapeServer DSSSD biases & leakage currents OK - see attachments 1 & 2 FEE64 temperatures OK - see attachment 3 DAQ run control OK - see attachment 4 Statistics - see attachments 5-16 N.B. some FEE64s are producing SYNCs, some not, in particular nnaida6 is not System-wide tests - see attachments 17-22 11.31 DAQ recovered with STOP-GO (no data transfer) AIDA SYNCs OK for all FEE64s 22.46 DAQ still running OK Power down in preparation for tomorrow's planned site-wide power outage