Message ID: 390     Entry time: Mon Oct 17 04:10:30 2016
Author: TD 
Subject: Monday 17 October 
12.05 DAQ found stopped
      Most recent file R49_226, last accessed 06.26
      Restart merger OK
      DAQ STOP OK

      System-wide checks - see attachments 1-2 
      SYNC errors and SYNC pulses fail

      ReSYNC OK
      System-wide checks OK
      Restart merger & tapeserver

      DSSSD bias & leakage currents OK - see attachments 3 & 4
      FEE64 temperatures OK - see attachment 5

      DAQ start
      #1 nnaida15 data transfer #1 tag drops at DAQ GO
         disable H & data transfer #1
         restart merger & tapeserver
         enable H & data transfer #1

12.22 #2 DAQ start (R51) OK

20.33 DAQ found stopped
      Most recent file R51_145, last accessed 18.30
      Restart merger OK
      DAQ STOP OK

      System-wide checks - see attachments 6 & 7
      SYNC errors and SYNC pulses fail

      DAQ RESET & SETUP OK
      Disable H & data transfer #1
      Restart merger & tapeserver
      Enable H & data transfer #1

20.41 DAQ GO (R52)
Attachment 1: 20.png  230 kB  Uploaded Mon Oct 17 05:12:00 2016  | Hide | Hide all
20.png
Attachment 2: 21.png  224 kB  Uploaded Mon Oct 17 05:12:00 2016  | Hide | Hide all
21.png
Attachment 3: 23.png  12 kB  Uploaded Mon Oct 17 05:12:08 2016  | Hide | Hide all
23.png
Attachment 4: 24.png  12 kB  Uploaded Mon Oct 17 05:12:08 2016  | Hide | Hide all
24.png
Attachment 5: 26.png  134 kB  Uploaded Mon Oct 17 05:13:10 2016  | Hide | Hide all
26.png
Attachment 6: 27.png  230 kB  Uploaded Mon Oct 17 13:36:18 2016  | Hide | Hide all
27.png
Attachment 7: 28.png  224 kB  Uploaded Mon Oct 17 13:36:18 2016  | Hide | Hide all
28.png
ELOG V3.1.4-unknown