AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
|
AIDA |
 |
|
Message ID: 4
Entry time: Thu Sep 11 16:12:06 2014
|
Author: |
Tom Davinson and photos by Patrick |
Subject: |
AIDA Tests at STFC DL - PCS, SLT, TD |
|
|
Changes from configuration described by Elog entry https://elog.ph.ed.ac.uk/AIDA/3
are as follows:
HV
Emco Q05-5 ISOLATED, PROPORTIONAL DC TO HV DC CONVERTER
Input TTi QL355T set +1.5V, Emco Q05-5 output +194.6V
Output RC filter 1uF x 1MOhm
i.e. a filtered, floating high voltage detector bias supply
cf. NIM-bin grounded Silena 7710 Quad Bias used previously
ASIC Adaptor PCB
nnaida11 & nnaida12 LK 3 & 7
nnaida13 & nnaida13 LK 1, 3, 5, & 7
HV + -> nnaida11 & nnaida12
HV - -> nnaida13 & nnaida14
DSSSD grounded locally at nnaida12 & nnaida13 adaptor PCBs
by ground links LK 1 & 5 (cf. Silena 7710 Quad Bias Supply
used previously).
ASIC settings
Default ASIC settings (see attachments 1-4)
Observe stable operation of both positive and negative polarity
FEE64s using default ASIC parameters and a floating HV detector
bias supply
Attachments
1-4 nnaida11, 12, 13, & 14 ASIC settings
5-8 nnaida11, 12, 13, & 14 hit/rate patterns
9-12 nnaida11, 12, 13, & 14 ASIC # 4 ADC spectra
13 nnaida12 ASIC # 1 waveforms
14 nnaida14 ASIC # 1 waveforms
15 Emco Q-series catalogue |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|