AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
AIDA
Draft saved at 00:00:00
Fields marked with
*
are required
Entry time:
Sat Mar 29 12:36:44 2025
Author
*
:
Subject
*
:
10.47 DAQ found stopped, no merger error messages Most recent file R78_281, last accessed 06.46 R78_281 ADC data synchronised DSSSD bias & leakage currents OK - see attachments 1 & 2 FEE64 temperatures OK - see attachment 3 System-wide checks - see attachments 4-11 Restart merger OK DAQ STOP OK DAQ RESET & SETUP OK ReSYNC OK System-wide checks OK Disable H & data transfer #1 Restart merger & tapeserver Enable H & data transfer #1 11.07 DAQ starts (R79) 18.39 DAQ OK System wide checks OK *except* SYNC errors & SYNC pulses - see attachments 12-15 18.47 DAQ stalls Most recent file R79_182, last accessed 18.46 System wide checks OK *except* SYNC errors & SYNC pulses - see attachments 16-19 18.56 DAQ starts (R80)