10.53 DAQ found stalled Most recent files R89_102 last accessed 07.17 DSSSD bias & leakage current OK - see attachments 1 & 2 FEE64 temperatures OK - see attachment 3 All system wide checks OK *except* Clock status, SYNC errors & SYNC pulses - see attachments 4-8 - unusual messages/reports One merger error messages - link 18 - see attachment 9 Restart merger DAQ STOP OK DAQ RESET & SETUP OK ReSYNC OK All system-wide checks OK Disable H & data transfer #1 Restart merger & tapeserver Enable H & data transfer #1 Standard soft reset did not work - one or more FEE64s dropped data transfer #1 link at DAQ GO Retry - one, more FEE64s indicate error state - run control tab unresponsive Power cycle FEE64s