AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
|
AIDA |
 |
|
Message ID: 411
Entry time: Mon Oct 31 02:00:01 2016
|
Author: |
TD |
Subject: |
Monday 31 October |
|
|
11.00 DAQ continues OK
DAQ STOP OK
Most recent files R91_568 last accessed 10.59
DSSSD bias & leakage current OK - see attachments 1 & 2
FEE64 temperatures OK - see attachment 3
System wide checks OK *except* SYNC pulses & errors - see attachments 4-7
14.20 Per FEE64 screenshots p+n junction and n+n ohmic strip waveforms - see attachments 8-11
Preamplifier output noise looks OK for all FEE64s *except* nnaida5 & nnaida6 (which
have significantly higher ADC and disc rates than other FEE64s)
21.10 DAQ found stalled
Most recent file R94_99 last accessed 18.35
Restart merger
DAQ STOP
System wide checks
DAQ RESET & SETUP
System-wide checks
Disable H & data transfer #1
Restart merger & tapeserver
Enable H & data transfer #1
21.16 DAQ start (R96) |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|