AIDA |
 |
|
Message ID: 448
Entry time: Wed Nov 9 22:16:11 2016
|
Author: |
CG, DK, TD, ML |
Subject: |
NP1412 R24 R25 |
|
|
07.15 R24 started. BRIKEN run 33.
76~78Ni beam setting.
F5 C target out. F7 CH2 in. F11 C out.
F11 plastic rate ~500cps.
1mm Al vairable degrader removed after ~3mins. Now have 4.4mm in.
07.26 R24 stopped for unknown reason - fill cycle had previously happened, similar scenario to yesterday.
Soft reset performed successfully.
07.36 R25 started.
No changes to beam settings above. F11 plastic rate ~500cps.
Offline implant rates:
*** DSSSD # 1 count: 894 old count: 682 dt: 1.02 s HEC rate: 207.66 Hz
*** DSSSD # 2 count: 781 old count: 590 dt: 1.02 s HEC rate: 187.09 Hz
*** DSSSD # 3 count: 900 old count: 682 dt: 1.02 s HEC rate: 213.54 Hz
*** DSSSD # 4 count: 790 old count: 594 dt: 1.02 s HEC rate: 191.99 Hz
*** DSSSD # 5 count: 421 old count: 317 dt: 1.02 s HEC rate: 101.87 Hz
*** DSSSD # 6 count: 162 old count: 128 dt: 1.02 s HEC rate: 33.30 Hz
08.13 Beam stopped.
corresponds to file R25_9.
08.15 Leakage currents attachments 1+2.
08.20 Takechi group changing settings: F5 C now in, F7 now empty, F11 C (6mm) now in.
Beam returned to AIDA.
Corresponding to 1.0Gb through file R25_10.
4.4mm of Al variable degrader currently in place. Will gradually remove.
08.23 All variable degrader removed.
Corresponds to start of file R25_11.
Rates from offline implant finder:
*** DSSSD # 1 count: 474 old count: 310 dt: 1.02 s HEC rate: 160.08 Hz
*** DSSSD # 2 count: 407 old count: 269 dt: 1.02 s HEC rate: 134.70 Hz
*** DSSSD # 3 count: 473 old count: 306 dt: 1.02 s HEC rate: 163.01 Hz
*** DSSSD # 4 count: 463 old count: 303 dt: 1.02 s HEC rate: 156.18 Hz
*** DSSSD # 5 count: 331 old count: 220 dt: 1.02 s HEC rate: 108.35 Hz
*** DSSSD # 6 count: 64 old count: 43 dt: 1.02 s HEC rate: 20.50 Hz
09.18 Beam stopped
Corresponds to file R25_23
09.20 Takechi et al.change settings - F5, F7, F11 out
Variable degrader 5.4mm Al in
First AIDA data file R25_24, BRIKEN 161110_0923_767778Ni_036
from file R25_26 (offline):
*** DSSSD # 1 count: 13306 old count: 13159 dt: 1.01 s HEC rate: 145.63 Hz
*** DSSSD # 2 count: 3578 old count: 3536 dt: 1.01 s HEC rate: 41.61 Hz
*** DSSSD # 3 count: 7699 old count: 7609 dt: 1.01 s HEC rate: 89.16 Hz
*** DSSSD # 4 count: 6191 old count: 6121 dt: 1.01 s HEC rate: 69.35 Hz
*** DSSSD # 5 count: 1891 old count: 1872 dt: 1.01 s HEC rate: 18.82 Hz
*** DSSSD # 6 count: 1377 old count: 1364 dt: 1.01 s HEC rate: 12.88 Hz
10.08 All system wide checks OK *except* nnaida8 & 16 fail clock status, nnaida23 fails ADC calibration
10.14 Beam stopped
Corresponds to file R25_36
10.19 Takechi et al.change settings - F5 in, F7 out, F11 in
Variable degrader all out
First AIDA data file R25_37, BRIKEN 161110_1016_767778Ni_37
10.26 Rate, Hitrate, 1.8.L and 1.8.W spectra - see attachments 3-12
All spectra cleared
10.30 Merge statistics - see attachments 13-17
10.32 Noted that nnaida19 histogram link has dropped - non-critical - continue DAQ
11.09 Change of target upstream: F5 empty, F7 CH2, F11 C 6mm
Corresponds to AIDA data file R25_48
11:14 Change completed
Corresponds to AIDA data file R25_49
11:21 All degraders out
Corresponds to AIDA data file R25_51
From AIDA R25_55:
*** DSSSD # 1 count: 241 old count: 0 dt: 1.02 s HEC rate: 235.75 Hz
*** DSSSD # 2 count: 210 old count: 0 dt: 1.02 s HEC rate: 205.42 Hz
*** DSSSD # 3 count: 242 old count: 0 dt: 1.02 s HEC rate: 236.72 Hz
*** DSSSD # 4 count: 209 old count: 0 dt: 1.02 s HEC rate: 204.44 Hz
*** DSSSD # 5 count: 118 old count: 0 dt: 1.02 s HEC rate: 115.43 Hz
*** DSSSD # 6 count: 46 old count: 0 dt: 1.02 s HEC rate: 45.00 Hz
11:40 lost beam
Current AIDA data R25_56
11:45 beam back
Current AIDA data R25_57
12.19 Takechi et al. change settings - F5 in, F7 out, F11 C 6mm
Corresponds to AIDA data file R25_65
All degraders 5.4mm Al in
12.24 All degraders out
Corresponds to AIDA data file R25_67
From offline analysis of R25_67
*** DSSSD # 1 count: 584 old count: 472 dt: 3.91 s HEC rate: 149.31 Hz
*** DSSSD # 2 count: 516 old count: 410 dt: 3.91 s HEC rate: 131.92 Hz
*** DSSSD # 3 count: 607 old count: 488 dt: 3.91 s HEC rate: 155.19 Hz
*** DSSSD # 4 count: 595 old count: 475 dt: 3.91 s HEC rate: 152.12 Hz
*** DSSSD # 5 count: 439 old count: 355 dt: 3.91 s HEC rate: 112.24 Hz
*** DSSSD # 6 count: 87 old count: 69 dt: 3.91 s HEC rate: 22.24 Hz
ADC & disc data synchronised - see attachments 17-18
13.38 R25_84 Bias and leakage currents ok - see attachments 19 and 20
13.57 FEE settings checked by saving current settings to file and comparing against the last saved settings. No change was found in the ASIC settings.
14.33 BRIKEN now on run 40 file name 161110_1419. DSSD Rates
*** DSSSD # 1 count: 138 old count: 0 dt: 1.02 s HEC rate: 135.54 Hz
*** DSSSD # 2 count: 119 old count: 0 dt: 1.02 s HEC rate: 116.88 Hz
*** DSSSD # 3 count: 139 old count: 0 dt: 1.02 s HEC rate: 136.52 Hz
*** DSSSD # 4 count: 134 old count: 0 dt: 1.02 s HEC rate: 131.61 Hz
*** DSSSD # 5 count: 101 old count: 0 dt: 1.02 s HEC rate: 99.20 Hz
*** DSSSD # 6 count: 20 old count: 0 dt: 1.02 s HEC rate: 19.64 Hz
15:14 BigRIPS now on Run 439 (CH2 Target in at F7)
AIDA file R25_104
15.17 BRIKEN now on run 41 file name 161110_1516
*** DSSSD # 1 count: 380 old count: 169 dt: 1.05 s HEC rate: 200.67 Hz
*** DSSSD # 2 count: 326 old count: 144 dt: 1.05 s HEC rate: 173.09 Hz
*** DSSSD # 3 count: 381 old count: 172 dt: 1.05 s HEC rate: 198.77 Hz
*** DSSSD # 4 count: 336 old count: 150 dt: 1.05 s HEC rate: 176.90 Hz
*** DSSSD # 5 count: 191 old count: 86 dt: 1.05 s HEC rate: 99.86 Hz
*** DSSSD # 6 count: 73 old count: 35 dt: 1.05 s HEC rate: 36.14 Hz
15.26 bias and leakage ok see attachment 21 and 22
16.04 BigRIPS Setting change now on 440 F5 In F7 Out F11 In
16.08 BRIKEN Now on run 42 File 161110_1608
16.09 Bias and leakage currents ok attachments 23 and 24
From offline r25_119
*** DSSSD # 1 count: 255 old count: 117 dt: 1.04 s HEC rate: 132.79 Hz
*** DSSSD # 2 count: 234 old count: 111 dt: 1.04 s HEC rate: 118.35 Hz
*** DSSSD # 3 count: 259 old count: 120 dt: 1.04 s HEC rate: 133.75 Hz
*** DSSSD # 4 count: 250 old count: 115 dt: 1.04 s HEC rate: 129.90 Hz
*** DSSSD # 5 count: 181 old count: 84 dt: 1.04 s HEC rate: 93.34 Hz
*** DSSSD # 6 count: 33 old count: 18 dt: 1.04 s HEC rate: 14.43 Hz
16.54 BRIKEN AND BigRIPS change runs. Corresponding to AIDA file R25_129
BigRIPS Now F5 In, F7 Out F11 Out
All variable Al degraders in 5.5mm
16.48 New BRIKEN run
Run number 043
File name: 161110_1658_767778Ni
17.02 Bias and leakage currents ok see attachment 25 and 26
DSSD Rates from offline file R25_131
*** DSSSD # 1 count: 269 old count: 127 dt: 1.04 s HEC rate: 136.32 Hz
*** DSSSD # 2 count: 224 old count: 107 dt: 1.04 s HEC rate: 112.32 Hz
*** DSSSD # 3 count: 257 old count: 118 dt: 1.04 s HEC rate: 133.44 Hz
*** DSSSD # 4 count: 160 old count: 71 dt: 1.04 s HEC rate: 85.44 Hz
*** DSSSD # 5 count: 29 old count: 13 dt: 1.04 s HEC rate: 15.36 Hz
*** DSSSD # 6 count: 3 old count: 2 dt: 1.04 s HEC rate: 0.96 Hz
17.48 BigRIPS change setting putting back the thick Carbon at F11 6mm. Corresponds to AIDA file R25_141
F5 In, F7 Out, F11 In C 6mm
All variable Al degraders removed.
BigRIPS run 442
17.52 BRIKEN starts run 44
File name 161110_1752767778Ni
AIDA file R25_142
bias and leakage currents ok attachments 27 and 28
DSSD rates offline from R25_142
*** DSSSD # 1 count: 275 old count: 136 dt: 1.03 s HEC rate: 134.63 Hz
*** DSSSD # 2 count: 240 old count: 114 dt: 1.03 s HEC rate: 122.04 Hz
*** DSSSD # 3 count: 275 old count: 134 dt: 1.03 s HEC rate: 136.57 Hz
*** DSSSD # 4 count: 266 old count: 130 dt: 1.03 s HEC rate: 131.73 Hz
*** DSSSD # 5 count: 180 old count: 86 dt: 1.03 s HEC rate: 91.05 Hz
*** DSSSD # 6 count: 46 old count: 24 dt: 1.03 s HEC rate: 21.31 Hz
18.16 System-wide checks OK *except* clock status (some/all of nnaida5,7,8,15&16 - see attachment 29) and ADC calibration (nnaida23)
18.24 DAQ stops OK
AIDA data filename R25_149
18.28 System wide checks show SYNC pulse and SYNC errors - see attachments 30-32
18.31 Stop & close merger & TapeServer
DAQ reset
DAQ setup - not complete - use Master Timestamp tab to ReSYNC
but note unusual(?) ReSYNC LSW value before second ReSYNC attempt- see attachments 33 & 34 before and after ReSYNC
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|