Entry time:
Wed Nov 23 01:47:28 2016
Author
*
:
Subject
*
:
10.45 DSSSD bias & leakage currents - see attachments 1 & 2 note - significant increase in leakage current (~0.8uA) since previous readings may be temperature transient or charging effect due to delta electrons due to BigRIPS tuning to F11 15cm Pb installed upstream of AIDA+BRIKEN some days ago FEE64 temperatures OK - see attachment 3 15.35 DSSD bias and leakage currents ok (attachments 4+5). Marginal increase over this morning. FEE temps ok (attachment 6).