AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  AIDA  ELOG logo
Message ID: 5     Entry time: Thu Sep 11 17:29:59 2014
Author: Tom Davinson 
Subject: AIDA Tests at STFC DL - To Do list 
In no particular order of priority:

1) Vary ASIC V_DD=3.3V within +/-10% operating range

2) Use 'clean' floating high voltage detector bias
    - check polarity
    - is there a difference between detector local
      ground on p+n junction or n+n ohmic sides of DSSSD? 
    - common or isolated detector local ground for multiple DSSSDs?
    - RFQ submitted for 1x CAEN N1419
    - additional filtering for Emco Q05-5?
    - other?

3) Backtrack changes to date
     - Kapton PCB mods
     - PSU set points
     - Standard FEE64 PSUs -> linearly regulated PSUs

4) Evaluate effect of ASIC parameters 'Ibias preamp SF' and 'Ibias preamp' on noise

5) Evaluate effect of ASIC parameter 'preAmp reference' on stability of positive and 
   negative input configurations

Comments and corrections welcome!