Message ID: 560     Entry time: Tue May 9 02:25:04 2017
Author: TD, OH 
Subject: Tuesday 9 May 
10.25 DAQ continues OK 
      Current file May2017/R1_79

      ASIC settings 2017Mar27-16.12.39 *except*
      slow comparator 0x64
      LEC/MEC fast comparator 0xff

      BNC PB-4 pulser OFF

      Detector biases & leakage currents OK - see attachments 1 & 2
      FEE64 Temperatures OK - see attachment 3
      'good events' statistics OK - see attachment 4
      No warnings/errors reported by merger since DAQ GO - see attachment 5

      *all* system wide checks OK

10.28 1.8.L ADC spectra - see attachments 6 & 7 
      1.8.W waveforms - see attachments 8-11

10.43 DAQ stopped 
      Current file May2017/R1_79

10.48 E17/F11 ambient temperature +24.5 deg C, RH 29.0%, d.p. +5.0 deg C
      Julabo FL11006 set point +20 deg C, actual +20.0 deg C, water level OK (c. 65%)


10.58 DAQ starts
      File May2017/R2 *IGNORE*

      Pulser walkthough 

      BNC PB-4 settings 
      Amplitude 90,000 - 10,000 @ 10,000 step
      x5 attenuator IN
      t_r 50ns, tau_d 50us
      polarity -
      rate ~250Hz 
      delay MIN, tail pulse

      ASIC settings 2017Mar27-16.12.39
      Slow comparator 0x0a
 
      See attachment 12

11.02 DAQ starts
      File May2017/R3 *IGNORE*

      Pulser walkthough 

      BNC PB-4 settings 
      Amplitude 12,000 - 2,000 @ 2,000 step
      x5 attenuator IN
      t_r 50ns, tau_d 50us
      polarity -
      rate ~250Hz 
      delay MIN, tail pulse

      ASIC settings 2017Mar27-16.12.39

      See attachment 13

Pulser walkthroughs R2 & R3 should be ignored as + polarity test input
had significantly lower amplitude (x2 oor more) than - polarity test input.
The + polarity test input is derived from tee'd output from BNC PB-4 and
EG&G Ortec 433 Sum & Invert - physically moving the cabling and connectors
between these two points changed tail pulse amplitude or produced a transient
(non-tail) pulse with duration ~ 100ns. Not clear which cable or connector
is the issue (all cables and connectors were individually replaced) but output
amplitude currently correct so will proceed.

11.22 DAQ starts
      File May2017/R4

      Pulser walkthough 

      BNC PB-4 settings 
      Amplitude 90,000 - 10,000 @ 10,000 step
      x5 attenuator IN
      t_r 50ns, tau_d 50us
      polarity -
      rate ~250Hz 
      delay MIN, tail pulse

      ASIC settings 2017Mar27-16.12.39
 
      See attachment 14

11.28 DAQ starts
      File May2017/R5

      Pulser walkthough 

      BNC PB-4 settings 
      Amplitude 12,000 - 2,000 @ 2,000 step
      x5 attenuator IN
      t_r 50ns, tau_d 50us
      polarity -
      rate ~250Hz 
      delay MIN, tail pulse

      ASIC settings 2017Mar27-16.12.39

      See attachment 15

17.25 E17/F11 ambient temperature +24.7 deg C, RH 28.1%, d.p. +5.0 deg C
      Julabo FL11006 set point +20 deg C, actual +20.0 deg C, water level OK (c. 65%)

17.32 Detector biases & leakage currents OK - see attachments 16-17
      FEE64 Temperatures OK - see attachment 18

17.40 DAQ starts
      File May2017/R6

      Continuing DAQ test

      BNC PB-4 settings 
      Amplitude 90,000
      x5 attenuator IN
      t_r 50ns, tau_d 50us
      polarity -
      frequency 25Hz range 
      delay MIN, tail pulse

      ASIC settings 2017Mar27-16.12.39
      slow comparator 0x20
Attachment 1: 41.png  12 kB  Uploaded Tue May 9 03:30:16 2017  | Hide | Hide all
41.png
Attachment 2: 42.png  12 kB  Uploaded Tue May 9 03:30:16 2017  | Hide | Hide all
42.png
Attachment 3: 43.png  136 kB  Uploaded Tue May 9 03:30:16 2017  | Hide | Hide all
43.png
Attachment 4: 44.png  90 kB  Uploaded Tue May 9 03:30:16 2017  | Hide | Hide all
44.png
Attachment 5: 45.png  294 kB  Uploaded Tue May 9 03:30:16 2017  | Hide | Hide all
45.png
Attachment 6: 46.png  120 kB  Uploaded Tue May 9 03:30:16 2017  | Hide | Hide all
46.png
Attachment 7: 47.png  119 kB  Uploaded Tue May 9 03:30:16 2017  | Hide | Hide all
47.png
Attachment 8: 48.png  141 kB  Uploaded Tue May 9 03:30:16 2017  | Hide | Hide all
48.png
Attachment 9: 49.png  158 kB  Uploaded Tue May 9 03:30:16 2017  | Hide | Hide all
49.png
Attachment 10: 50.png  131 kB  Uploaded Tue May 9 03:30:16 2017  | Hide | Hide all
50.png
Attachment 11: 51.png  144 kB  Uploaded Tue May 9 03:30:16 2017  | Hide | Hide all
51.png
Attachment 12: 60.png  111 kB  Uploaded Tue May 9 04:20:05 2017  | Hide | Hide all
60.png
Attachment 13: 61.png  113 kB  Uploaded Tue May 9 04:20:05 2017  | Hide | Hide all
61.png
Attachment 14: 63.png  112 kB  Uploaded Tue May 9 04:31:46 2017  | Hide | Hide all
63.png
Attachment 15: 64.png  112 kB  Uploaded Tue May 9 04:31:46 2017  | Hide | Hide all
64.png
Attachment 16: 70.png  12 kB  Uploaded Tue May 9 10:35:43 2017  | Hide | Hide all
70.png
Attachment 17: 71.png  12 kB  Uploaded Tue May 9 10:35:43 2017  | Hide | Hide all
71.png
Attachment 18: 72.png  135 kB  Uploaded Tue May 9 10:35:43 2017  | Hide | Hide all
72.png
ELOG V3.1.4-unknown