Entry time:
Sat Mar 29 12:37:11 2025
Author
*
:
Subject
*
:
11.19 DAQ continues OK Current file May2017/R10_346 ASIC settings 2017Mar27-16.12.39 *except* slow comparator 0x0a LEC/MEC fast comparator 0xff Detector biases & leakage currents OK - see attachments 1 & 2 FEE64 Temperatures OK - see attachment 3 'good events' statistics OK - see attachment 4 No warnings/errors reported by merger since DAQ GO - see attachment 5 *all* system wide checks OK 11.22 Analysis of data file R10_246 - see attachment 6 dead-time/idle-time < 1% AIDA SYNC100 variations ~700/97000 10.10 E17/F11 ambient temperature +24.5 deg C, RH 38.0%, d.p. +9.4 deg C Julabo FL11006 set point +20 deg C, actual +20.0 deg C, water level OK (c. 65%) 11.23 DAQ stops OK Current file May2017/R10_347 Files May2017/R10* *deleted* ASIC settings 2017Mar27-16.12.39 *except* slow comparator 0x0a LEC/MEC fast comparator 0xff BNC PB-4 settings Amplitude 90,000 x5 attenuator IN t_r 50ns, tau_d 50us polarity - rate 250Hz range delay MIN, tail pulse Check ASIC Control x2 Zero all spectra 11.30 DAQ starts File May2017/R11 18.18 Detector biases & leakage currents OK - see attachments 7 & 8 FEE64 Temperatures OK - see attachment 9 18.24 E17/F11 ambient temperature +24.5 deg C, RH 36.8%, d.p. +8.7 deg C Julabo FL11006 set point +20 deg C, actual +20.0 deg C, water level OK (c. 65%)