Subject
*
:
11.11 DAQ found stopped Most recent file May2017/R11_245 (last accessed 06:09) Detector biases & leakage currents OK - see attachments 1 & 2 FEE64 Temperatures OK - see attachment 3 'good events' statistics zero - see attachment 4 Merger and Tapeserver terminal server windows are gone - see attachment 5 *all* system wide checks OK No relevant system log messages (/var/log/messages) c. 06.00 DAQ STOP attempt #1 did not stop #2 timeout nnaida1, nnaida2-24 did not stop #3 timeout nnaida1, nnaida2-24 did not stop Restart, configure and setup/go merger terminal server Restart TapeServer terminal server DAQ STOP attempt #3 all stop OK 11.24 Analysis of data file R11_245 - see attachment 6 dead-time/idle-time < 0.5% AIDA SYNC100 variations ~160/78000 11.34 E17/F11 ambient temperature +24.5 deg C, RH 37.0%, d.p. +9.0 deg C Julabo FL11006 set point +20 deg C, actual +20.0 deg C, water level OK (c. 65%) 13.06 Check ASIC Control x2 Zero all spectra Restart terminal servers for Httpd for Merger/TapeServer, Merger & TapeServer Restart browser for Merger & TapeServer N.B. *no FEE64 power cycle, no DAQ RESET, SETUP etc* 11.30 DAQ starts File May2017/R12 13.10 DAQ GO - data transfer links for nnaida11 & nnaida12 dropped DAQ STOP Disable all data transfer links Restart terminal servers for Httpd for Merger/TapeServer, Merger & TapeServer Restart browser for Merger & TapeServer Enable all data transfer links DAQ GO OK File May2017/R13 ASIC settings 2017Mar27-16.12.39 *except* slow comparator 0x0a LEC/MEC fast comparator 0xff
Encoding
:
HTML
ELCode
plain
Suppress Email notification
Attachment 1:
Drop attachments here...
Draft saved at 00:00:00
ELOG V3.1.4-unknown