AIDA |
 |
|
Message ID: 604
Entry time: Sun May 28 02:16:04 2017
|
Author: |
TD, CB |
Subject: |
Sunday 28 May |
|
|
10.15 DAQ continues OK
Current file May2017/R31_239
Detector biases; leakage currents OK - see attachments 1 & 2
FEE64 Temperatures OK - see attachment 3
good events; statistics OK - see attachment 4
No warnings/errors reported by merger since DAQ GO - see attachment 5
*all* system wide checks OK *except* ADC calibration of nnaida1, 21 and nnaida1 clock status (status=6)
10.30 E17/F11 ambient temperature +23.8 deg C, RH 40.9%, d.p. +9.6 deg C
Julabo FL11006 set point +20 deg C, actual +20.0 deg C, water level OK (c. 95% - refilled)
10.35 Layouts 1, 3, 4, 7 and 8 - see attachments 6-13
Note - as expected, zero HEC events, layouts 5 & 6 not shown
10.58 DAQ stop - transient timeout nnaida1 - followed by successful stop
File May2017/R31_248
Delete R31
11.12 Disable data transfer
DAQ GO - transient timeout nnaida1 - followed by successful start
nnaida1 system console - see attachment 14
12:00 Moved plastic veto scintillator downstream in an attempt to reduce noise
Rates in noisy FEEs reduced by 20-30%
14:30 DAQ continues OK
Found that pulser induces counts in some HEC (nnaida20, 14 in particular). See attachment 15. Pulser off
removes these events completely.
These events could be mistaken for implantation events during data acquisition.
Increasing the fast comparator threshold HEC from 2 -> 3 suppresses the events. Increasing the threshold
to 4 does not appear to have further effects.
Even with the increased threshold some HEC (e.g. 1.0H) still see the pulser. See attachment 16 (fast
comparator threshold HEC was 0x3).
Checked data is synchronised in file R32_33
15:29 Set the threshold back to 2. We will perform this cut off-line in the analysis. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|