AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  AIDA  ELOG logo
Message ID: 627     Entry time: Sat Jun 3 05:32:09 2017
Author: TD 
Subject: Report - high - MBS correlation scaler anomalies  
The analysis of files RIBF127R1/R1_1478 to R1_1486 is appended. This sequence of files should
encompass a correlation reset request (from BRIKEN to AIDA)

There appear to be a number of anomalies

1) MBS correlation scalers observed with one missing 16-bit data value (bits 16-31 or bits 0-15)
   across approximately five files (~20 minutes)

2) The first (presumably) MBS correlation scaler following reset corresponds to a time ~300s (25MHz clock)   
Attachment 1: testmbs  98 kB  Uploaded Sat Jun 3 11:31:03 2017  | Show | Show all
ELOG V3.1.4-unknown