AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
AIDA
Draft saved at 00:00:00
Fields marked with
*
are required
Entry time:
Sat Mar 29 12:32:00 2025
Author
*
:
Subject
*
:
> 0:04 Biases and leak currents shown as attachment #1, 12JunBias1.png > System-wide checks: nnaida13 fails ADC calib, all others okay > Temperatures shown as attachment #2, 12JunTemp1.png > Implants are 20 to 40 cps, mostly in SSD2 and SSD6 see attachment #3 12JunImplant1.png > Statistics shown attachment #4, 12JunStats1.png > Analysis shown as R9_175.txt attachment #5 > > > 00:17 No implants in AIDA nor proper evt. numbers in BRIKEN R9_177, expect beam is down > Operators call, RF dropped of IRC > > 00:23 Problem of another RF. > > 00:24 Beam is back R9_179 for AIDA > > 00:36 AIDA R9_182 at run close. > > 00:39 AIDA R9_183, BRIKEN 141, run BigRIPS 4031 > > 00:50 We stop the beam to adjust degraders in F11. AIDA R9_186 > BRIKEN and BigRIPS runs stopped > > 00:54 Removed 1.0 mm Al, added 0.5 mm (net is subtracting 0.5 mm Al) > R9_187 AIDA run, BRIKEN #142, BigRIPS #4032 > > 01:34 AIDA R9_198 at changeover > > 01:35 Same AIDA run R9_198, BRIKEN #143, BigRIPS #4033 > System wide checks okay (except ADC calib on nnaida13) > > 02:34 R9_214 at changeover. > > 2:36 R9_215 with new runs BRIKEN 144 and BigRIPS 4034. > More implants are now seen in SSD3 rather than SSD2. Zeroed MIDAS histos, so updated later with stats. > Biases as attachment #6, 12JunBiases2.png > > 02:45 Beam stopped R9_217. Beam stripper has a problem. > > 02:48 Beam is back, R9_220 > > 03:12 R9_224 operators call, beam is stopped from stripper foil issue > > 03:18 Beam is back, AIDA run R9_226 > > 03:34 R9_230 at changeover > System wide checks all okay, except nnaida13 ADC calib as above. > > 03:35 The end of R9_230 corresponds with BRIKEN 145 and BigRIPS 4034. > > 03:51 Operators call, beam looks stopped. R9_235 at AIDA. Change stripper foil > > 03:57 R9_236 when beam returns. > > 04:35 We noticed beam is gone. R9_246 for AIDA. > > 04:36 Beam comes back. R9_247 BRIKEN 146 and BigRIPS 4035 > Attachment #7 12JunImplant2.png shows implantation profile, confirming SSD 3 is the main one with new > degrader setting > Statistics shown as attachment #8, 12JunStats2.png > Analysis shown as attachment #9, R9_244.txt > No ts timewarps, no MBS timewarps, deadtime on nnaida13,21 nearly 1%, most others small. > > 05:20 Noticed /data10 only has ~250 GB left, so we should do something to resolve it. > First thing is, not much compression is done recently. So I begin to compress /TapeData/RIBF128/R7* with > nice 10 > However, compression at this rate will take much too long so it looks like something should be removed. > Confirmed that RIBF127R1 data file names and sizes are the same between /TapeData and the new external HDD > at /ntfs > > 05:33 AIDA R9_262, BRIKEN 147, BigRIPS 4036 > > 06:34 AIDA R9_279 at run change > > 06:36 BRIKEN run 148, BigRIPS 4037 > System-wide checks passed, except nnaida13 ADC calib still fails. > > 06:54 R9_284 beam is stopped. > > 6:59 Beam is back. > > 07:33 R9_295 at start of BRIKEN 149, BigRIPS 4037 > > 07:36 No syncs are seen at BRIKEN (but they were seen at the start of other runs) > Implant profiles shown as Attachment #10, 12JunImplants3.png > Several resets or new runs on BRIKEN / BigRIPS are done to troubleshoot > > 07:48 AIDA passes all system checks (besides the nnaida13 ADC calib) > It seems (part of) the issue is triggers from BigRIPS. > > 08:04 BigRIPS scalers show events are presented, but no triggers are accepted. > Some part of the DAQ is vetoing the events basically. > > 08:11 Baba san came on shift. Issue was ccnet03 crashed. We reset the Camac crate and reinitialized, DAQ comes back > AIDA R9_306, BRIKEN #152, BigRIPS # 4041
Encoding
:
HTML
ELCode
plain
Suppress Email notification
Attachment 1:
Drop attachments here...
Draft saved at 00:00:00
ELOG V3.1.4-unknown