AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
|
AIDA |
 |
|
Message ID: 668
Entry time: Sun Jun 18 02:56:43 2017
|
Author: |
TD |
Subject: |
Sunday 18 June |
|
|
10.56 Detector biases & leakage currents OK - see attachments 1 & 2
USB-controlled ac mains relay input #1 fuse blown - replaced
11.12 *all* system wide checks OK *except* nnaida21 fails ADC calibration
BNC PB-4
Amplitude 90,000
x5 attenuator IN
t_r 100ns tau_d 50us
- polarity
frequency 25Hz range 2.6Hz
delay MIN, tail pulse, INT ref
ASIC settings 2017Mar27-16.12.39
11.36 nnaida13 & nnaida11 1*W spectra - see attachments 3 & 4
per FEE64 1.8.W spectra - see attachments 5 & 6
good events statistics - see attachment 7
Note c. 3kHz modulated HF noise has disappeared overnight
12.00 2x unused (possibly faulty) MACBs removed from AIDA NIM bins preparatory to UK return
ambient temperature +24.0 deg C, d.p. +8.9 deg C, RH 38.8%
Julabo FL11006 set point +20.0 deg C, actual +20.0 deg C, water level c. 75%
14.03 nnaida13 1*W spectra
Ibias LF feedback 0x8 (default) - see attachments 8 & 10
Ibias LF feedback 0xf - see attachments 9 & 11
14.30 BRIKEN 3He and clovers off
no obvious change in preamp waveform noise
14.33 DSSSD bias off
14.37 FEE64 power off
14.43 2x NIM bins off
BNC PB-4 pulser off
USB-controlled ac mains relay off, interlock enabled, USB cable removed
ac mains circuit off
Tektronix AFG3201C off
Julabo FL11006 off
3ph supply off |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|