Entry  Wed Apr 22 03:33:01 2015, TD, FEE64 Fast Discriminator OR o/p via MACB tek00000.tiftek00001.tif
    Reply  Thu Apr 23 09:02:57 2015, TD, Patrick, FEE64 Fast Discriminator OR o/p via MACB 
Message ID: 68     Entry time: Thu Apr 23 09:02:57 2015     In reply to: 64
Author: TD, Patrick 
Subject: FEE64 Fast Discriminator OR o/p via MACB 
> Illustration of the trigger timing of the AIDA FEE64 fast discriminator OR output via the
> MACB front panel Trigger outputs.
> 
> Channel 1: BNC PB-4 Pulser Trigger Output
> Channel 2: BNC PB-4 Pulser Analogue Output
> Channel 3: MACB Trigger Output
> 
> MACB Trigger Output: NIM complement
>                      width ~ 7us
>                      delay ~ 750ns (with respect to BNC PB_4 Pulser Analogue Output)
> 
> Example illustrates MAC Trigger Output from nnaida5 with standard ASIC settings and
> LEC/MEC fast discriminator 0x40

It would be useful to show the Trigger signal at the different levels of the MACB distribution system... then 
the difference will show the delay through the MACB layer.

Are you able to estimate the length of cable between the pulser and the adaptor boards.

What are the conclusions about the delay .... is it too much to be useful ?