> 11.27 DAQ continues > Not writing to disk > > BNC PB-4 > Amplitude 90,000 > x5 attenuator IN > t_r 100ns tau_d 50us > - polarity > frequency range 25Hz > delay MIN, tail pulse, INT ref > > ASIC settings 2017Oct23-14.38.03 > *all* FEE64s slow comparators 0xa > > Detector biases & leakage currents OK - see attachments 1 & 2 > FEE64 Temperatures OK - see attachment 3 > good events statistics stable & OK - see attachments 4 > *all* system wide checks OK *except* nnaida21 fails ADC calibration > > 11.47 Ambient temperature +23.9 deg C, d.p. +9.0 deg C, RH 39.0% > > 16:45 System wide checks all ok > FEE temps ok - Attachment 5 > Stats - Attachment 6 > Bias and leakage ok - Attachment 7