AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
AIDA
Draft saved at 00:00:00
Fields marked with
*
are required
Entry time:
Sat Mar 29 12:26:40 2025
Author
*
:
Subject
*
:
> 09.56 DAQ stopped OK yesterday > > BNC PB-4 > Amplitude 90,000 > x5 attenuator IN > t_r 100ns tau_d 50us > - polarity > frequency range 25Hz > delay MIN, tail pulse, INT ref > > ASIC settings 2017Oct23-14.38.03 > *all* FEE64s slow comparators 0xa *except* nnaida13 0xe, nnaida19-22 0xc > > Detector biases & leakage currents OK - see attachments 1 & 2 > FEE64 Temperatures OK - see attachment 3 > *all* system wide checks OK *except* nnaida21 fails ADC calibration > Julabo FL11006 recirculating chiller coolant reservoir refilled - set point +20 deg C, actual +20.0 deg C > Ambient temperature +23.6 deg C, d.p. +6.2 deg C, RH 32.2% > > AIDA withdrawn from (CH2)n matrix > > DAQ started - no data transfer > good events statistics stable - see attachment 4 > *all* system wide checks OK *except* nnaida21 fails ADC calibration > > 13:30 System wide checks ok > Stats - Attachment 6 > FEE temperatures ok - Attachment 7 > leakage and bias ok - Attachment 8 > > 17.15 Detector biases & leakage currents OK - see attachments 9 & 10 > FEE64 Temperatures OK - see attachment 11 > good events statistics stable - see attachment 12 > *all* system wide checks OK *except* nnaida21 fails ADC calibration > Ambient temperature +25.2 deg C, d.p. +5.7 deg C, RH 28.5%