11.33 DAQ running OK - date transfer not enabled BNC PB-4 Amplitude 90,000 x5 attenuator IN t_r 100ns tau_d 50us - polarity frequency range 25Hz delay MIN, tail pulse, INT ref ASIC settings 2017Oct23-14.38.03 *all* FEE64s slow comparators 0xa *except* nnaida13 0xe, nnaida19-22 0xc Detector biases & leakage currents OK - see attachments 1 & 2 FEE64 Temperatures OK - see attachment 3 DAQ statistics OK - see attachment 4 *all* system wide checks OK *except* nnaida21 fails ADC calibration Julabo FL11006 recirculating chiller set point +20 deg C, actual +20.0 deg C Ambient temperature +26.2 deg C, d.p. +6.2 deg C, RH 28.3% 14.40 Detector biases & leakage currents OK - see attachments 5 & 6 FEE64 Temperatures OK - see attachment 7 DAQ statistics OK - see attachment 8 *all* system wide checks OK *except* nnaida21 fails ADC calibration 14.50 DAQ starts AIDA data file Oct17/R18 14.53 per FEE64 rates (20us/channel) - see attachment 9 No indication of c. 400us period noise as indicated by DAQ statistics attachment 8 15.46 DAQ stopped DAQ start - data tranfser disabled 19.40 Detector biases & leakage currents OK - see attachments 10 & 11 FEE64 Temperatures OK - see attachment 12 DAQ statistics OK - see attachment 13 *all* system wide checks OK *except* nnaida21 fails ADC calibration