Entry time:
Sat Mar 29 11:17:33 2025
Author
*
:
Subject
*
:
> 14.16 DAQ running OK - data transfer not enabled > > BNC PB-4 > Amplitude 90,000 > x5 attenuator IN > t_r 100ns tau_d 50us > - polarity > frequency range 25Hz > delay MIN, tail pulse, INT ref > > ASIC settings 2017Oct23-14.38.03 > *all* FEE64s slow comparators 0xa *except* nnaida13 0xe, nnaida19-22 0xc > > Detector biases & leakage currents OK - see attachments 1 & 2 > FEE64 Temperatures OK - see attachment 3 > DAQ statistics OK - see attachment 4 > *all* system wide checks OK *except* nnaida21 fails ADC calibration > Julabo FL11006 recirculating chiller set point +20 deg C, actual +20.0 deg C > Ambient temperature +23.9 deg C, d.p. +10.8 deg C, RH 43.2% > > 15.05 per FEE64 Rate, 1.8.L and 1.8.W spectra - note common x & y scales > Peak width nnaida5 63 ch FWHM > Peak width nnaida8 82 ch FWHM > > 15.09 All histograms zero'd > > 18.25 Detector biases & leakage currents OK - see attachments 11 & 12 > FEE64 Temperatures OK - see attachment 13 > DAQ statistics OK - see attachment 14 > *all* system wide checks OK *except* nnaida21 fails ADC calibration