AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  AIDA  ELOG logo
Message ID: 692     Entry time: Tue Oct 31 00:34:10 2017
Author: TD, OH 
Subject: Tuesday 31 October 
09.34 DAQ running OK - data transfer not enabled

      BNC PB-4
      Amplitude 90,000
      x5 attenuator IN
      t_r 100ns tau_d 50us
      - polarity
      frequency range 25Hz
      delay MIN, tail pulse, INT ref

      ASIC settings 2017Oct23-14.38.03
      *all* FEE64s slow comparators 0xa *except* nnaida13 0xe, nnaida19-22 0xc

      Detector biases & leakage currents OK - see attachments 1 & 2
      FEE64 Temperatures OK - see attachment 3
      DAQ statistics OK - see attachment 4
      *all* system wide checks OK *except* nnaida21 fails ADC calibration

15.02 Detector biases & leakage currents OK - see attachments 5 & 6
      FEE64 Temperatures OK - see attachment 7
      DAQ statistics OK - see attachment 8
      *all* system wide checks OK *except* nnaida21 fails ADC calibration
      Ambient temperature +24.3 deg C, d.p. -0.2 deg C, RH 19.8%
      Julabo FL11006 recirculating chiller set point +20 deg C, actual +20.0 deg C, water level OK c. mid-point
middle-upper marker
     
Attachment 1: 20.png  12 kB  | Hide | Hide all
20.png
Attachment 2: 21.png  12 kB  | Hide | Hide all
21.png
Attachment 3: 22.png  140 kB  | Hide | Hide all
22.png
Attachment 4: 23.png  90 kB  | Hide | Hide all
23.png
Attachment 5: 200.png  12 kB  Uploaded Tue Oct 31 06:04:09 2017  | Hide | Hide all
200.png
Attachment 6: 201.png  12 kB  Uploaded Tue Oct 31 06:04:09 2017  | Hide | Hide all
201.png
Attachment 7: 202.png  136 kB  Uploaded Tue Oct 31 06:04:09 2017  | Hide | Hide all
202.png
Attachment 8: 203.png  90 kB  Uploaded Tue Oct 31 06:04:09 2017  | Hide | Hide all
203.png
ELOG V3.1.4-unknown