AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  AIDA  ELOG logo
Message ID: 694     Entry time: Wed Nov 1 02:24:20 2017
Author: OH 
Subject: Wednesday 1 November 
11:24 DAQ running OK - data transfer not enabled

      BNC PB-4
      Amplitude 90,000
      x5 attenuator IN
      t_r 100ns tau_d 50us
      - polarity
      frequency range 25Hz
      delay MIN, tail pulse, INT ref

      ASIC settings 2017Oct23-14.38.03
      *all* FEE64s slow comparators 0xa *except* nnaida13 0xe, nnaida19-22 0xc

      Detector biases & leakage currents OK - see attachments 1
      FEE64 Temperatures OK - see attachment 2
      DAQ statistics OK - see attachment 3
      *all* system wide checks OK *except* nnaida21 fails ADC calibration
Attachment 1: 011117_1120_bias.png  19 kB  | Hide | Hide all
011117_1120_bias.png
Attachment 2: 011117_1122_feetemps.png  107 kB  | Hide | Hide all
011117_1122_feetemps.png
Attachment 3: 011117_1123_stats.png  57 kB  | Hide | Hide all
011117_1123_stats.png
ELOG V3.1.4-unknown