Entry time:
Sat Mar 29 11:14:52 2025
Author
*
:
Subject
*
:
> 08.09 DAQ running OK - data transfer not enabled > > BNC PB-4 > Amplitude 90,000 > x5 attenuator IN > t_r 100ns tau_d 50us > - polarity > frequency range 25Hz > delay MIN, tail pulse, INT ref > > ASIC settings 2017Oct23-14.38.03 > *all* FEE64s slow comparators 0xa *except* nnaida19-22 0xc > > Detector biases & leakage currents OK - see attachments 1 & 2 > FEE64 Temperatures OK - see attachment 3 > DAQ statistics OK - see attachment 4 > *all* system wide checks OK *except* nnaida21 fails ADC calibration - see attachment 6 > PLL and DCM Clock Lock Monitor OK - see attachment 5 > > 08.22 per FEE64 Rate, 1.8.L and 1.8.W spectra - see attachments 7-12 > nnaida3 78 ch FWHM > nnaida17 66 ch FWHM > > 16.43 Detector biases & leakage currents OK - see attachments 13 & 14 > FEE64 Temperatures OK - see attachment 15 > DAQ statistics OK - see attachment 16 > *all* system wide checks OK *except* nnaida21 fails ADC calibration - see attachment 6 > PLL and DCM Clock Lock Monitor OK - see attachment 5 > > per FEE64 Rate, 1.8.L pectra - see attachments 17-20 > nnaida3 78 ch FWHM > nnaida17 68 ch FWHM