AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
|
AIDA |
 |
|
Message ID: 700
Entry time: Fri Nov 17 10:03:25 2017
|
Author: |
CB, TD |
Subject: |
Friday 17 November |
|
|
19.03 DAQ started OK - data transfer not enabled
BNC PB-4
Amplitude 90,000
x5 attenuator IN
t_r 100ns tau_d 50us
- polarity
frequency range 25Hz
delay MIN, tail pulse, INT ref
ASIC settings 2017Oct23-14.38.03
*all* FEE64s slow comparators 0xa
Detector biases & leakage currents OK - see attachments 1
FEE64 Temperatures OK - see attachment 2
DAQ statistics OK - see attachment 3
*all* system wide checks OK *except* nnaida7 fails ADC calibration
Rate look OK. A few dropped channels. See attach 4
Waveforms look OK. Noise at reasonable levels. See attach 5 & 6
19:15 Started run Nov17/R1
Kbyte/sec written 2611
Merger data rate 346210 items/sec
19:26 Observed increase in rate in asics nnaida13, 19 (>50 000)
Noise profile looks typical. See waveforms - attach 7,8
Increased FEE64 slow comparators for nnaida 13, 19 from 0xa to 0xc
Rate decreases significantly. See attach 9
19.40 Beam on
BigRIPS 115Nb setting
WAS3ABI + YSO
Good Event, merger and data rates OK - see attachments 10-12
8x FEE64s <10k
8x FEE64s 10-20k
6x FEE64s 20-30k
2x FEE64s 30-40k
19.57 per FEE64 1.8.W spectra (2ms FSR) - see attachments 14-15
20.06 per FEE64 1.8.L spectra - see attachments 16-17
nnaida17 67 ch FWHM
nnaida8 73 ch FWHM
20.50 DAQ starts
RIKEN run: 005
Wasabi Run: 4048
BigRIPS run: 2057
AIDA run R2_0
Listmode DLT file:171117_2050_115Nb_005.dlt
BRIKEN Rate (h:12): ~ 42 cps
F11 Rate (h:18): ~ 37 cps
CLOVERS Rate (h:15): ~ 33 cps
AIDA Veto Plastic Rate (h:19): ~ 23
WAS3ABi in
YSO removed
AIDA inserted to HPGe clover cryostat position
BigRIPS 115Nb setting
Degrader 5mm Al
BNC PB-4
Amplitude 90,000
x5 attenuator IN
t_r 100ns tau_d 50us
- polarity
frequency range 25Hz
delay MIN, tail pulse, INT ref
ASIC settings 2017Oct23-14.38.03
*all* FEE64s slow comparators 0xa *except* nnaida13 0xf, nnaida19 0xc, nnaida20-21 0xb
20.57 all spectra zero'd
Good Event stats - see attachment 18
7x FEE64s <10k
5x FEE64s 10-20k
4x FEE64s 20-30k
4x FEE64s 30-40k
3x FEE64s 40-50k
1x FEE64 >50k (nnaida13 73k)
21.22 Not all ASICs settings restored from 2017Oct23-14.38.03
Some p+n FEE64s Ibias LF feedback 0x8 *not* 0xf as expected - now changed manually
AIDA data file Nov17/R2_5
ASIC settings saved to 2017Nov17-21.24.47
Updated options
21.33 New BigRIPS, BRIKEN, WAS3ABi files started
AIDA continues from R2_6
Date: 17/11/017
Start: 2133
RIKEN run: 006
Wasabi Run: 4061
BigRIPS run: 2058
AIDA run R2_06
Listmode DLT file:171117_2132_115Nb_006.dlt
BRIKEN Rate (h:12): ~ 42 cps
F11 Rate (h:18): ~ 37 cps
CLOVERS Rate (h:15): ~ 33 cps
AIDA Veto Plastic Rate (h:19): ~ 23
21.40 Detector biases & leakage currents OK - see attachments 19 & 20
FEE64 Temperatures OK - see attachment 21
DAQ statistics OK - see attachment 22
*all* system wide checks OK *except* nnaida7 fails ADC calibration
per FEE64 1.8.* spectra (20ms FSR) see attachments 23 & 24
p+n spectra now look OK with Ibias LF feedback changed from 0x8 to 0xf
note 3kHz appears in nnaida19, 21 & 22
22.41 DAQ stop
AIDA data file Nov17/R2_15
*use R2_8 - R2_15* for offline analysis
22.50 AIDA withdrawn from BRIKEN
YSO re-installed
23.10 Detector biases & leakage currents OK - see attachments 25 & 26
FEE64 Temperatures OK - see attachment 27
DAQ statistics OK - see attachment 28
7x FEE64s <10k
4x FEE64s 10-20k
8x FEE64s 20-30k
2x FEE64s 30-40k
2x FEE64s 40-50k
1x FEE64 >50k (nnaida13 64k)
*all* system wide checks OK *except* nnaida7 fails ADC calibration
per FEE64 1.8.W spectra (20ms FSR) - see attachments 29 & 30
per FEE64 rate spectra linear and log scales - see attachments 31 & 32
23.30 all spectra zero'd |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|