Message ID: 701     Entry time: Sat Nov 18 03:37:46 2017
Author: TD 
Subject: Offline analysis RIBF128/R5_124-R5_148, Oct17/R17_31-R17_34 and Nov17/R2_8-R2_16 data 
Offline analysis of

AIDA RIBF128/R5_124 to R5_148 data "130Ag setting", June 2017 RIBF128 experiment         *yellow*
AIDA Oct17/R17_31 to R17_34 data   "84Ge setting" , October 2017, 78Ni experiment        *red*
AIDA Nov17/R2_8 to R2_16 data      "115Nb setting", November 2017 RIBF139 experiment     *blue*

N.B. 
1) no software thresholds
2) no PID correlation
3) unless otherwise stated, all implant energies are the stopping layer implant energies

Attachment 1 - *total* implant energy for AIDA DSSSD stack (5.6MeV/channel)

Attachment 2 - implant energy for stopping DSSSD layer (20MeV/channel)

Attachments 3-4 - per pixel implant-decay correlation times (163.84us/ch)

The 130Ag data (yellow) clearly shows the "20ms bump" but the later test data (red & blue)
with the revised AIDA FEE64 adaptor PCBs *and* using ASIC setting I_bias LF feedback 0x8 -> 0xf
for all p+n FEE64s shows (with the limited statistics available) no evidence of a "20ms bump".
Attachment 1: 100.png  29 kB  | Hide | Hide all
100.png
Attachment 2: 101.png  27 kB  | Hide | Hide all
101.png
Attachment 3: 102.png  29 kB  | Hide | Hide all
102.png
Attachment 4: 103.png  26 kB  | Hide | Hide all
103.png