Author
*
:
Subject
*
:
> 09.27 Detector biases & leakage currents OK - see attachments 1 & 2 > FEE64 Temperatures OK - see attachment 3 > DAQ statistics OK - see attachment 4 > 8x FEE64s <10k > 2x FEE64s 10-20k > 5x FEE64s 20-30k > 3x FEE64s 30-40k > 1x FEE64s 40-50k > 5x FEE64 >50k (nnaida13 255k) > *all* system wide checks OK *except* nnaida7 fails ADC calibration > > 09.42 per FEE64 rate, 1.8.L, 1.8.W (20ms FSR) spectra - see attachments 5-10 > all spectra zero'd > > 17.05 Detector biases & leakage currents OK - see attachments 11 & 12 > FEE64 Temperatures OK - see attachment 13 > DAQ statistics OK - see attachment 14 > 7x FEE64s <10k > 4x FEE64s 10-20k > 2x FEE64s 20-30k > 4x FEE64s 30-40k > 2x FEE64s 40-50k > 5x FEE64 >50k (nnaida13 261k) > *all* system wide checks OK *except* nnaida7 fails ADC calibration > > per FEE64 rate, 1.8.L, 1.8.W (20ms FSR) spectra - see attachments 15-20 > pulser peak width nnaida17 69 ch FWHM, nnaida8 72 ch FWHM