AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  AIDA  ELOG logo
Message ID: 708     Entry time: Thu Nov 23 00:41:43 2017
Author: TD, CB 
Subject: Thursday 23 November  
09.33 Detector biases & leakage currents OK - see attachments 1 & 2
      FEE64 Temperatures OK - see attachment 3
      DAQ statistics OK - see attachment 4
      7x FEE64s <10k
      4x FEE64s 10-20k
      6x FEE64s 20-30k
      2x FEE64s 30-40k
      0x FEE64s 40-50k
      5x FEE64 >50k (nnaida13 280k)
      *all* system wide checks OK *except* nnaida7 fails ADC calibration

      per FEE64 rate, 1.8.L, 1.8.W (20ms FSR) spectra - see attachments 5-10
      pulser peak width nnaida17 70 ch FWHM, nnaida8 72 ch FWHM
Attachment 1: 1.png  12 kB  | Hide | Hide all
1.png
Attachment 2: 2.png  12 kB  | Hide | Hide all
2.png
Attachment 3: 3.png  136 kB  | Hide | Hide all
3.png
Attachment 4: 4.png  91 kB  | Hide | Hide all
4.png
Attachment 5: 11.png  145 kB  | Hide | Hide all
11.png
Attachment 6: 12.png  157 kB  | Hide | Hide all
12.png
Attachment 7: 13.png  122 kB  | Hide | Hide all
13.png
Attachment 8: 14.png  124 kB  | Hide | Hide all
14.png
Attachment 9: 17.png  170 kB  | Hide | Hide all
17.png
Attachment 10: 18.png  155 kB  | Hide | Hide all
18.png
ELOG V3.1.4-unknown