Message ID: 708
Entry time: Thu Nov 23 00:41:43 2017
|
Author: |
TD, CB |
Subject: |
Thursday 23 November |
|
|
09.33 Detector biases & leakage currents OK - see attachments 1 & 2
FEE64 Temperatures OK - see attachment 3
DAQ statistics OK - see attachment 4
7x FEE64s <10k
4x FEE64s 10-20k
6x FEE64s 20-30k
2x FEE64s 30-40k
0x FEE64s 40-50k
5x FEE64 >50k (nnaida13 280k)
*all* system wide checks OK *except* nnaida7 fails ADC calibration
per FEE64 rate, 1.8.L, 1.8.W (20ms FSR) spectra - see attachments 5-10
pulser peak width nnaida17 70 ch FWHM, nnaida8 72 ch FWHM |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|