> 11:05 Bias and leakage currents ok - attachment 1 > FEE temperatures ok > System wide checks ok > Currently on file R9_74 > Temperature 25.3C, Humidity 42.9%, D.P +12.1C > > 13:27 Bias and leakage currents ok - attachment 2 > FEE Temperatures ok > System wide checks ok > > Analysis of the pulser run R9_52 -> R9_74 has shown no major losses in terms of apparent broken bond wires. > The large gap in FEE64 29 can be explained by an ASIC dropping out: See https://elog.ph.ed.ac.uk/AIDA/730 > Another odd trend appears in the channels relating to DSSD 8 on the x side. - attachment 5 > The pattern shows one strip on one strip off for the first 32 strips. > nnaida18 is also the FEE which is running at a much higher rate during this run. - attachment 3 > I'm not sure that I believe that these channels are actually relate to broken bond wires. > > Here are the hit patterns for the 8 DSSD - attachment 4 > Attached is also the x and y histograms for DSSD 8 which shows a weird pattern - attachment 5